On April 30, 2021, Ashton, Gage P.; Charsley, Edward L.; Harding, Lindsay P.; Parkes, Gareth M. B. published an article.SDS of cas: 118-55-8 The title of the article was DSC temperature calibration over the range -125°C to 50°C. And the article contained the following:
The lack of Nationally Certified Reference Materials for the calibration of DSC equipment in the sub-ambient range has been addressed by the introduction of a Certified Reference Material of purified cyclohexane by the National Metrol. Institute of Japan. This sample, in conjunction with the CRM material Ph salicylate and the ITS-90 fixed point material mercury, has enable a DSC-thermomicroscopy system to be accurately calibrated over the temperature range -90°C to 50°C. The DSC has then been used to measure values for the solid-solid transitions in the potential Certified Reference Materials ammonium sulfate, adamantane and ammonium dihydrogen phosphate, which have the advantage of being thermally stable and easy to handle, and for the m.p. of di-Ph ether. In addition, the effect of temperature cycling on these materials has been investigated. The experimental process involved the reaction of Phenyl Salicylate(cas: 118-55-8).SDS of cas: 118-55-8
The Article related to differential scanning calorimetry temperature calibration, Thermodynamics, Thermochemistry, and Thermal Properties: Calorimetry, Thermal Analysis, Thermogravimetry and other aspects.SDS of cas: 118-55-8